Calibration network based on alanine/ESR dosimetry.
Appl. Radiat. Isot. 44, 23-31 (1993)
It is shown that an international network of ESR dosimetry laboratories can exchange alanine samples for precise interlaboratory calibration in the absorbed dose range 0.01-100 kCy. For the experimental verification the alanine/ESR dosimetry laboratories and calibration centers of both the CSF-Forschungszentrum fur Umwelt und Gesundheit, Neuherberg, and the Istituto Superiore di Sanita (ISS), Rome, were mutually acting as a partial network. A blind test, together with the National Physical Laboratory (NPL), Teddington, UK, revealed agreement between the NPL administered doses and the doses reported by GSF and ISS within ±1%. | It is shown that an international network of ESR dosimetry laboratories can exchange alanine samples for precise interlaboratory calibration in the absorbed dose range 0.01-100 kGy. For the experimental verification the alanine/ESR dosimetry laboratories and calibration centers of both the GSF-Forschungszentrum fur Umwelt und Gesundheit, Neuherberg, and the Istituto Superiore di Sanita (ISS), Rome, were mutually acting as a partial network. A blind test, together with the National Physical Laboratory (NPL), Teddington, UK, revealed agreement between the NPL administered doses and the doses reported by GSF and ISS within ±1%.
Altmetric
Weitere Metriken?
Zusatzinfos bearbeiten
[➜Einloggen]
Publikationstyp
Artikel: Journalartikel
Dokumenttyp
Wissenschaftlicher Artikel
Schlagwörter
Dosimetry; transfer dosimetry; alanine dosimetry; ESR dosimetry; free radical dosimetry; calibration network
ISSN (print) / ISBN
0969-8043
e-ISSN
1872-9800
Zeitschrift
Applied Radiation and Isotopes
Quellenangaben
Band: 44,
Heft: 1-2,
Seiten: 23-31
Verlag
Elsevier
Begutachtungsstatus
Peer reviewed
Institut(e)
Institute of Radiation Protection (ISS)