Calibration Network based on Alanine/ESR Dosimetry?.

Appl. Radiat. Isot. 44, 23-31 (1993)
DOI
Open Access Green möglich sobald Postprint bei der ZB eingereicht worden ist.
It is shown that an international network of ESR dosimetry laboratories can exchange alanine samples for precise interlaboratory calibration in the absorbed dose range 0.01–100 kGy. For the experimental verification the alanine/ESR dosimetry laboratories and calibration centers of both the GSF-Forschungszentrum fur Umwelt und Gesundheit, Neuherberg, and the Istituto Superiore di Sanità (ISS), Rome, were mutually acting as a partial network. A blind test, together with the National Physical Laboratory (NPL), Teddington, UK, revealed agreement between the NPL administered doses and the doses reported by GSF and ISS within ±1%.
Altmetric
Weitere Metriken?
Zusatzinfos bearbeiten [➜Einloggen]
Publikationstyp Artikel: Journalartikel
Dokumenttyp Wissenschaftlicher Artikel
Korrespondenzautor
Schlagwörter Dosimetry; transfer dosimetry; alanine dosimetry; ESR dosimetry; free radical dosimetry; calibration network
ISSN (print) / ISBN 0969-8043
e-ISSN 1872-9800
Quellenangaben Band: 44, Heft: 1-2, Seiten: 23-31 Artikelnummer: , Supplement: ,
Verlag Elsevier
Nichtpatentliteratur Publikationen
Begutachtungsstatus Peer reviewed