Open Access Green möglich sobald Postprint bei der ZB eingereicht worden ist.
Reproducibility study of TLD-100 micro-cubes at radiotherapy dose level.
Appl. Radiat. Isot. 50, 573-577 (1999)
The precision of the thermoluminescent response of Harshaw micro-cube dosimeters (TLD-100), evaluated in both Harshaw thermoluminescent readers 5500 and 3500, for 1 Gy dose value, was investigated. The mean reproducibility for micro-cubes, pre-readout annealed at 100°C for 15 min, evaluated with the manual planchet reader 3500, is 0.61% (1 standard deviation). When micro-cubes are evaluated with the automated hot-gas reader 5500, reproducibility values are undoubtedly worse, mean reproducibility for numerically stabilised dosimeters being equal to 3.27% (1 standard deviation). These results indicate that the reader model 5500, or, at least, the instrument used for the present measurements, is not adequate for micro-cube evaluation, if precise and accurate dosimetry is required. The difference in precision is apparently due to geometry inconsistencies in the orientation of the imperfect micro-cube faces during readout, requiring careful and manual reproducible arrangement of the selected micro-cube faces in contact with the manual reader planchet.
Altmetric
Weitere Metriken?
Zusatzinfos bearbeiten
[➜Einloggen]
Publikationstyp
Artikel: Journalartikel
Dokumenttyp
Wissenschaftlicher Artikel
ISSN (print) / ISBN
0969-8043
e-ISSN
1872-9800
Zeitschrift
Applied Radiation and Isotopes
Quellenangaben
Band: 50,
Heft: 3,
Seiten: 573-577
Verlag
Elsevier
Begutachtungsstatus
Peer reviewed
Institut(e)
Institute of Radiation Protection (ISS)