Gómez-Ros, J.M.* ; de Carlan, L.* ; Franck, D.* ; Gualdrini, G.* ; Lis, M.* ; López, M.A.* ; Moraleda, M.* ; Zankl, M.

Analysis of a computational problem involving complex voxel geometries.

Radiat. Prot. Dosim. 131, 24-27 (2008)
DOI
Open Access Green möglich sobald Postprint bei der ZB eingereicht worden ist.
This communication briefly summarises the results obtained from the 'International comparison on MC modeling for in vivo measurement of Americium in a knee phantom' organised within the EU Coordination Action CONRAD (Coordinated Network for Radiation Dosimetry) as a joint initiative of EURADOS working groups 6 (computational dosimetry) and 7 (internal dosimetry). Monte Carlo simulations using the knee voxel phantom proved to be a viable approach to provide the calibration factor needed for in vivo measurements.