PuSH - Publikationsserver des Helmholtz Zentrums München

Naserke, H.E.* ; Dozio, N.* ; Ziegler, A.-G.* ; Bonifacio, E.*

Comparison of a novel micro-assay for insulin autoantibodies with the conventional radiobinding assay.

Diabetologia 41, 681-683 (1998)
DOI PMC
Open Access Green möglich sobald Postprint bei der ZB eingereicht worden ist.
Measurement of insulin autoantibodies (IAA) with a novel micro radiobinding assay which requires only 20 microl of serum was compared with that in a conventional radiobinding assay which uses 600 microl of serum. IAA were measured with both assays in samples from 94 new onset insulin-dependent diabetes mellitus (IDDM) patients, 97 control subjects. and 48 first degree relatives of IDDM patients selected for having IAA in the conventional radiobinding assay. Overall, 227 (95%) of 239 samples tested were concordant. and IAA levels correlated well (r2 = 0.7) between the two assays. Discordant results were obtained in 7 new onset patients, 4 control subjects, and 1 first degree relative, and these had low IAA levels in the respective assays. Sensitivity and specificity in the new onset IDDM patients and control subjects were 69% and 98% for the micro radiobinding assay and 72% and 98% for the conventional radiobinding assay. The use of the micro radiobinding assay should greatly facilitate islet related antibody screening, particularly in children.
Altmetric
Weitere Metriken?
Zusatzinfos bearbeiten [➜Einloggen]
Publikationstyp Artikel: Journalartikel
Dokumenttyp Wissenschaftlicher Artikel
ISSN (print) / ISBN 0012-186X
e-ISSN 1432-0428
Zeitschrift Diabetologia
Quellenangaben Band: 41, Heft: 6, Seiten: 681-683 Artikelnummer: , Supplement: ,
Verlag Springer
Verlagsort Berlin ; Heidelberg [u.a.]
Begutachtungsstatus Peer reviewed
Institut(e) Institute of Diabetes Research (IDF)
Institute of Pancreatic Islet Research (IPI)