PuSH - Publikationsserver des Helmholtz Zentrums München

Conductivity of alanine solution for high level dosimetry.

Radiat. Prot. Dosim. 47, 585-587 (1993)
DOI
Open Access Green möglich sobald Postprint bei der ZB eingereicht worden ist.
The amino acid alanine is well known as a dosimetric detector material for high level dosimetry. Its application is based on the formation of radicals by ionising radiation. The free radicals are earlier detected by electron spin resonance (ESR) spectroscopy or chemically after dissolving the irradiated samples. Of all these methods the ESR/alanine system is the most advanced and is suggested for reference dosimetry. At present, however, the high cost of the system is a serious handicap for a large scale routine application in radiation plants. In this study the variation of electrical conductivity of L-alanine solution with applied dose is investigated in the range from 0.5-200 kGy. The conductivity was measured with a 50 MHz RF oscillator. This readout method is uncomplicated and may be suitable for routine application. The experiments were performed with L-alanine solution in glass ampoules.
Altmetric
Weitere Metriken?
Zusatzinfos bearbeiten [➜Einloggen]
Publikationstyp Artikel: Journalartikel
Dokumenttyp Wissenschaftlicher Artikel
Korrespondenzautor
ISSN (print) / ISBN 0144-8420
e-ISSN 1742-3406
Quellenangaben Band: 47, Heft: 1-4, Seiten: 585-587 Artikelnummer: , Supplement: ,
Verlag Oxford University Press
Verlagsort Oxford
Nichtpatentliteratur Publikationen
Begutachtungsstatus Peer reviewed