PuSH - Publikationsserver des Helmholtz Zentrums München

In-phantom 10B capture rates for medical applications at a reactor therapy facility.

Radiat. Prot. Dosim. 44, 433-436 (1992)
DOI
Open Access Green möglich sobald Postprint bei der ZB eingereicht worden ist.
Methods and results are described for the determination of 10B reaction rates in a simple homogeneous phantom of head size by experiment and Monte Carlo calculation. A depth-rate matrix for monoenergetic neutrons is established. Neutron spectra are folded into the matrix in order to simulate experimental results from an existing reactor neutron therapy facility and to predict the characteristics of a facility which is under design.
Impact Factor
Scopus SNIP
Altmetric
0.000
0.000
Tags
Anmerkungen
Besondere Publikation
Auf Hompepage verbergern

Zusatzinfos bearbeiten
Eigene Tags bearbeiten
Privat
Eigene Anmerkung bearbeiten
Privat
Auf Publikationslisten für
Homepage nicht anzeigen
Als besondere Publikation
markieren
Publikationstyp Artikel: Journalartikel
Dokumenttyp Wissenschaftlicher Artikel
Sprache englisch
Veröffentlichungsjahr 1992
HGF-Berichtsjahr 0
ISSN (print) / ISBN 0144-8420
e-ISSN 1742-3406
Quellenangaben Band: 44, Heft: 1-4, Seiten: 433-436 Artikelnummer: , Supplement: ,
Verlag Oxford University Press
Verlagsort Oxford
Begutachtungsstatus Peer reviewed
Scopus ID 0026480382
Erfassungsdatum 1992-12-31