PuSH - Publikationsserver des Helmholtz Zentrums München

Evaluation of an automatic blood pressure device for use in blood pressure screening programs.

Methods Inf. Med. 23, 75-81 (1984)
DOI
Open Access Green möglich sobald Postprint bei der ZB eingereicht worden ist.
An automatic blood pressure recorder (Bosch Tonotest) was evaluated on a subsample of the study population of the MuniCh Blood Pressure Study I (MBS I). The purpose of this evaluation was to test an automatic device for its usability in screening programs like the Munich Blood Pressure Program (MBP). For all subjects in MBS I three blood pressure measurements were taken using the random-zero sphygmomanometer. For a subs ample of subjects, in addition to the three blood pressure measurements with the random-zero sphygmomanometer, two measurements were made, one again with the random-zero sphygmomanometer and one with the automatic device. These last measurements were made in alternating sequence (cross-over design). The automatic device tends to underestimate the systolic blood pressure while the diastolic blood pressure readings agree well. The ability to identify people with high blood pressure was worse for the automatic device. The differences between the automatic device and the random-zero sphygmomanometer in sensitivity, specificity, error rates and their implications for a screening program are discussed.
Altmetric
Weitere Metriken?
Zusatzinfos bearbeiten [➜Einloggen]
Publikationstyp Artikel: Journalartikel
Dokumenttyp Wissenschaftlicher Artikel
Korrespondenzautor
Schlagwörter Screening Tests, Evaluation of Automatic Blood Pressure Recorders, Predictive Value
ISSN (print) / ISBN 0026-1270
e-ISSN 2511-705X
Quellenangaben Band: 23, Heft: 2, Seiten: 75-81 Artikelnummer: , Supplement: ,
Verlag Schattauer
Nichtpatentliteratur Publikationen
Begutachtungsstatus Peer reviewed
Institut(e) Department for Medical Information Systems (MEDIS)