Dose conversion coefficients for partial-fan CBCT scans.
Proc. SPIE 10132:1013241 (2017)
Due to the increasing number of cone-beam CT (CBCT) devices on the market, reliable estimates of patient doses for these imaging modality is desired. Cone-beam CT devices differ from conventional CT not only by a larger collimation but also by different recording modes. In this work, it has been investigated whether reliable patient doses can be obtained for CBCT devices in partial-fan mode using pre-computed slices. As an exemplary case, chest CBCT scans for the ICRP reference adult models has been examined. By normalizing organ doses to CTDI100w , the resulting dose conversion coefficients for CBCT could be well reproduced by precomputed slices, with a relative difference in the effective dose conversion coefficients of less than 10%.
Impact Factor
Scopus SNIP
Web of Science
Times Cited
Scopus
Cited By
Altmetric
Publikationstyp
Artikel: Journalartikel
Dokumenttyp
Wissenschaftlicher Artikel
Typ der Hochschulschrift
Herausgeber
Schlagwörter
Keywords plus
Sprache
englisch
Veröffentlichungsjahr
2017
Prepublished im Jahr
HGF-Berichtsjahr
2017
ISSN (print) / ISBN
0277-786X
e-ISSN
1996-756X
ISBN
Bandtitel
Konferenztitel
Medical Imaging 2017: Physics of Medical Imaging
Konferzenzdatum
11 February 2017
Konferenzort
Orlando, Florida, United States
Konferenzband
Quellenangaben
Band: 10132,
Heft: ,
Seiten: ,
Artikelnummer: 1013241
Supplement: ,
Reihe
Verlag
SPIE
Verlagsort
Tag d. mündl. Prüfung
0000-00-00
Betreuer
Gutachter
Prüfer
Topic
Hochschule
Hochschulort
Fakultät
Veröffentlichungsdatum
0000-00-00
Anmeldedatum
0000-00-00
Anmelder/Inhaber
weitere Inhaber
Anmeldeland
Priorität
Begutachtungsstatus
Peer reviewed
POF Topic(s)
30203 - Molecular Targets and Therapies
30504 - Mechanisms of Genetic and Environmental Influences on Health and Disease
Forschungsfeld(er)
Radiation Sciences
PSP-Element(e)
G-501391-001
G-501100-008
Förderungen
Copyright
Erfassungsdatum
2017-03-20