PuSH - Publikationsserver des Helmholtz Zentrums München

Dose conversion coefficients for partial-fan CBCT scans.

Proc. SPIE 10132:1013241 (2017)
Postprint DOI
Open Access Green
Due to the increasing number of cone-beam CT (CBCT) devices on the market, reliable estimates of patient doses for these imaging modality is desired. Cone-beam CT devices differ from conventional CT not only by a larger collimation but also by different recording modes. In this work, it has been investigated whether reliable patient doses can be obtained for CBCT devices in partial-fan mode using pre-computed slices. As an exemplary case, chest CBCT scans for the ICRP reference adult models has been examined. By normalizing organ doses to CTDI100w , the resulting dose conversion coefficients for CBCT could be well reproduced by precomputed slices, with a relative difference in the effective dose conversion coefficients of less than 10%.
Altmetric
Weitere Metriken?
Zusatzinfos bearbeiten [➜Einloggen]
Publikationstyp Artikel: Journalartikel
Dokumenttyp Wissenschaftlicher Artikel
Korrespondenzautor
ISSN (print) / ISBN 0277-786X
e-ISSN 1996-756X
Konferenztitel Medical Imaging 2017: Physics of Medical Imaging
Konferzenzdatum 11 February 2017
Konferenzort Orlando, Florida, United States
Zeitschrift Proceedings of SPIE
Quellenangaben Band: 10132, Heft: , Seiten: , Artikelnummer: 1013241 Supplement: ,
Verlag SPIE
Nichtpatentliteratur Publikationen
Begutachtungsstatus Peer reviewed