Open Access Green möglich sobald Postprint bei der ZB eingereicht worden ist.
A BeOSL finger ring dosemeter.
Radiat. Meas. 131:106234 (2020)
The Individual Monitoring Service (IMS) at the Helmholtz Zentrum Miinchen together with Dosimetrics GmbH has developed and characterized a new ring dosemeter for photon radiation for the measurement of H-p(0,07), based on BeOSL technology. This work describes the new finger ring dosemeter and provides data from Monte Carlo simulation and from the radiological characterization of the dosemeters according to IEC 62387, such as the energy and angular response functions of the dosemeter, its dose linearity and the coefficient of variation. Furthermore, the opacity of the dosemeter enclosure and light conditions for packing and unpacking of the detectors are addressed.
Altmetric
Weitere Metriken?
Zusatzinfos bearbeiten
[➜Einloggen]
Publikationstyp
Artikel: Journalartikel
Dokumenttyp
Wissenschaftlicher Artikel
Schlagwörter
Extremity Dosimetry ; Finger Ring ; Osl ; Beo ; Monte Carlo Simulation ; Beosl; System
ISSN (print) / ISBN
1350-4487
e-ISSN
1879-0925
Zeitschrift
Radiation Measurements
Quellenangaben
Band: 131,
Artikelnummer: 106234
Verlag
Elsevier
Verlagsort
The Boulevard, Langford Lane, Kidlington, Oxford Ox5 1gb, England
Begutachtungsstatus
Peer reviewed
Institut(e)
Auswertungsstelle (AWST)