PuSH - Publikationsserver des Helmholtz Zentrums München

Hödlmoser, H. ; Greiter, M. ; Bandalo, V. ; Brönner, J. ; Haninger, T. ; Emmerl, M. ; Mende, E. ; Scheubert, P.* ; Esser, R.* ; Figel, M.

A BeOSL finger ring dosemeter.

Radiat. Meas. 131:106234 (2020)
DOI
Open Access Green möglich sobald Postprint bei der ZB eingereicht worden ist.
The Individual Monitoring Service (IMS) at the Helmholtz Zentrum Miinchen together with Dosimetrics GmbH has developed and characterized a new ring dosemeter for photon radiation for the measurement of H-p(0,07), based on BeOSL technology. This work describes the new finger ring dosemeter and provides data from Monte Carlo simulation and from the radiological characterization of the dosemeters according to IEC 62387, such as the energy and angular response functions of the dosemeter, its dose linearity and the coefficient of variation. Furthermore, the opacity of the dosemeter enclosure and light conditions for packing and unpacking of the detectors are addressed.
Altmetric
Weitere Metriken?
Zusatzinfos bearbeiten [➜Einloggen]
Publikationstyp Artikel: Journalartikel
Dokumenttyp Wissenschaftlicher Artikel
Korrespondenzautor
Schlagwörter Extremity Dosimetry ; Finger Ring ; Osl ; Beo ; Monte Carlo Simulation ; Beosl; System
ISSN (print) / ISBN 1350-4487
e-ISSN 1879-0925
Quellenangaben Band: 131, Heft: , Seiten: , Artikelnummer: 106234 Supplement: ,
Verlag Elsevier
Verlagsort The Boulevard, Langford Lane, Kidlington, Oxford Ox5 1gb, England
Nichtpatentliteratur Publikationen
Begutachtungsstatus Peer reviewed