PuSH - Publikationsserver des Helmholtz Zentrums München

Dragan, G.-C. ; Kohlmeier, V. ; Orasche, J. ; Schnelle-Kreis, J. ; Forbes, P.B.C.* ; Breuer, D.* ; Zimmermann, R.

Development of a personal aerosol sampler for monitoring the particle-vapour fractionation of SVOCs in workplaces.

Ann. Work Expo. Health 64, 903-908 (2020)
Postprint DOI PMC
Open Access Green
Semi-volatile organic compounds (SVOCs), partitioned between particulates and vapours of an aerosol, require special attention. The toxicological effects caused by the inhalation of such aerosols may depend on the concentration and in which phase the organic compounds are found. A personal denuder-gas-particle separation aerosol sampler was developed to provide information about the partitioning of aerosols from organic compounds. The sampler was tested in a series of controlled laboratory experiments, which confirmed the capability and accuracy of the sampler to measure gas-particle mixtures. An average difference of 14.8 ± 4.8% was found between sampler and reference laboratory instruments. The obtained results showed that our sampler enables a more accurate measurement of the SVOC aerosols' gas-particle fractionation, compared with that of conventional samplers.
Altmetric
Weitere Metriken?
Zusatzinfos bearbeiten [➜Einloggen]
Publikationstyp Artikel: Journalartikel
Dokumenttyp Wissenschaftlicher Artikel
Korrespondenzautor
Schlagwörter Denuder ; Gas–particle Partitioning ; Personal Sampling ; Semi-volatile Organic Compound (svoc); Silicone-rubber Traps; Occupational-exposure; Semivolatile; Transmission; Evaporation; Collection; Denuders; Design; Mass
ISSN (print) / ISBN 2398-7308
e-ISSN 2398-7316
Quellenangaben Band: 64, Heft: 8, Seiten: 903-908 Artikelnummer: , Supplement: ,
Verlag Oxford University Press
Verlagsort Oxford
Nichtpatentliteratur Publikationen
Begutachtungsstatus Peer reviewed
Förderungen German Social Accident Insurance (DGUV)