Detailed comparison of Xenon APPI (9.6/8.4 eV), Krypton APPI (10.6/10.0 eV), APCI, and APLI (266 nm) for gas chromatography high resolution mass spectrometry of standards and complex mixtures.
J. Am. Soc. Mass Spectrom. 34, 1632-1646 (2023)
Photoionization schemes for mass spectrometry, either by laser or discharge lamps, have been widely examined and deployed. In this work, the ionization characteristics of a xenon discharge lamp (Xe-APPI, 9.6/8.4 eV) have been studied and compared to established ionization schemes, such as atmospheric pressure chemical ionization, atmospheric pressure photoionization with a krypton discharge lamp (Kr-APPI, 10.6/10 eV) and atmospheric pressure laser ionization (266 nm). Addressing the gas-phase ionization behavior has been realized by gas chromatography coupled to high-resolution mass spectrometry without the usage of a dopant. For standard substances, it has been found that Xe-APPI is able to ionize a broad range of polycyclic aromatic hydrocarbons as well as their heteroatom-containing and alkylated derivatives. However, thiol and ester compounds could not be detected. Moreover, Xe-APPI revealed a high tendency to generate oxygenated artifacts, most likely due to a VUV absorption band of oxygen at 148 nm. Beneficially, almost no chemical background, commonly caused by APCI or Kr-APPI due to column blood, plasticizers or impurities, is observed. This advantage is noteworthy for evolved gas analysis without preseparation or for chromatographic coelution. For the complex mixtures, Xe-APPI revealed the predominant generation of radical cations via direct photoionization with a high selectivity toward aromatic core structures with low alkylation. Interestingly, both Xe-APPI and Kr-APPI could sensitively detect sterane cycloalkanes, validated by gas chromatographic retention. The narrowly ionized chemical space could let Xe-APPI find niche applications, e.g., for strongly contaminated samples to reduce the background.
Altmetric
Weitere Metriken?
Zusatzinfos bearbeiten
[➜Einloggen]
Publikationstyp
Artikel: Journalartikel
Dokumenttyp
Wissenschaftlicher Artikel
Schlagwörter
Atmospheric Pressure Ionization ; Gas Chromatography ; Mass Spectrometry ; Photoionization ; Polycyclic Aromatic Hydrocarbons; Atmospheric-pressure Photoionization; Polycyclic Aromatic-hydrocarbons; Chemical-ionization; Electrospray-ionization; Laser Ionization; Thermal-degradation; Selective Detection; Ion-source; Petroleum; Ms
e-ISSN
1044-0305
Quellenangaben
Band: 34,
Heft: 8,
Seiten: 1632-1646
Verlag
Elsevier
Verlagsort
1155 16th St, Nw, Washington, Dc 20036 Usa
Nichtpatentliteratur
Publikationen
Begutachtungsstatus
Peer reviewed
Förderungen
German Research Foundation (DFG)
Horizon 2020 program
Horizon 2020 program