Kocijancic, M.* ; Roche, J.* ; Peter, A. ; Hörber, S.

Evaluation of the sysmex CN-6000 coagulation analyzer for routine and specialized coagulation testing in a central laboratory.

Clin. Lab. 71, 493-501 (2025)
DOI
Open Access Green möglich sobald Postprint bei der ZB eingereicht worden ist.
Background: The Sysmex CN-6000 is a fully automated high-throughput coagulation analyzer. The objective of this study was to evaluate the analytical performance of the analyzer for routine and special coagulation testing in a high-throughput central laboratory of a university hospital. Methods: The within- and between-day precision and accuracy of 29 coagulation parameters were evaluated on the Sysmex CN-6000 using commercially available quality control materials. Patient plasma samples were used to compare results of coagulation measurements between the Sysmex CN-6000 and the Atellica COAG 360, including plasma samples with visual interference. The sample throughputs of both analyzers were compared using plasma samples from healthy volunteers. Results: Within- and between-day coefficients of variation were acceptable for all assays tested on the Sysmex CN6000. High correlation and good agreement were observed when comparing coagulation results from the Sysmex CN-6000 and the Atellica COAG 360. Samples with visual interference showed comparable coagulation results between the two analyzers, with slightly better detection by the Sysmex CN-6000. The sample throughput per hour for analysis of a panel of five coagulation parameters was higher with the Sysmex CN-6000 compared to the Atellica COAG 360 (247 vs. 193 tests). Conclusions: The Sysmex CN-6000 demonstrated excellent analytical performance for a large number of coagulation parameters and has a high throughput capacity, ideal for the needs of a central laboratory with a high volume of routine and specialized coagulation testing. (Clin. Lab. 2025;71:493-501. DOI: 10.7754/Clin.Lab.2024.240922)
Impact Factor
Scopus SNIP
Altmetric
0.600
0.000
Tags
Anmerkungen
Besondere Publikation
Auf Hompepage verbergern

Zusatzinfos bearbeiten
Eigene Tags bearbeiten
Privat
Eigene Anmerkung bearbeiten
Privat
Auf Publikationslisten für
Homepage nicht anzeigen
Als besondere Publikation
markieren
Publikationstyp Artikel: Journalartikel
Dokumenttyp Wissenschaftlicher Artikel
Schlagwörter hemostasis; coagulation; core laboratory; throughput; HIL check; Hemostasis
Sprache englisch
Veröffentlichungsjahr 2025
HGF-Berichtsjahr 2025
ISSN (print) / ISBN 1433-6510
Zeitschrift Clinical Laboratory
Quellenangaben Band: 71, Heft: 3, Seiten: 493-501 Artikelnummer: , Supplement: ,
Verlag Clinical Laboratory Publ.
Verlagsort Mainz
Begutachtungsstatus Peer reviewed
POF Topic(s) 90000 - German Center for Diabetes Research
Forschungsfeld(er) Helmholtz Diabetes Center
PSP-Element(e) G-502400-001
Erfassungsdatum 2025-05-09