CTDIvol: A suitable normalization for CT dose conversion coefficients at different tube voltages?
In: Proceedings (SPIE Medical Imaging 2012: Physics of Medical Imaging, 04. - 09.02.12, San Diego, USA). Bellingham, USA: SPIE, 2012.:83131P (Proc. SPIE ; 8313)
DOI
Open Access Green as soon as Postprint is submitted to ZB.
Reliable estimates for patient doses for CT examinations are desirable for the patients themselves as well as for new epidemiological studies. It has been shown that dose conversion coefficients normalized to CTDIvol provide rather scanner independent quantities. In this work, it is demonstrated that this normalization provides also tube voltage independent values by simulating axial CT scans of a seven-year old infant and an adult. The differences in the effective dose conversion coefficients per CTDIvol between 80 and 120 kV is for most body heights below 5%. Only at the height of the testes and the thyroid the difference can be as large as 15%. This results in differences of the effective dose conversion coefficient per CTDIvol between 80 and 120 kV of less than 6-7% for typical CT examinations.
Altmetric
Additional Metrics?
Publication type
Article: Conference contribution
Document type
Thesis type
Corresponding Author
Keywords
CTDvol; CT examinations; patient dose; conversion coefficients; effective dose
Keywords plus
ISSN (print) / ISBN
0277-786X
e-ISSN
1996-756X
ISBN
9780819489623
Book Volume Title
Conference Title
SPIE Medical Imaging 2012: Physics of Medical Imaging, 04. - 09.02.12, San Diego, USA
Conference Date
Conference Location
Proceedings Title
Proceedings
Quellenangaben
Volume: 8313,
Issue: ,
Pages: ,
Article Number: 83131P
Supplement: ,
Series
Proceedings of SPIE
Publisher
SPIE
Publishing Place
Bellingham, USA
University
University place
Faculty
Publication date
0000-00-00
Application date
0000-00-00
Patent owner
Further owners
Application country
Patent priority
Reviewing status
Peer reviewed
Grants
Copyright