PuSH - Publication Server of Helmholtz Zentrum München

Dosymetry in the Bremsstrahlung Beam of an 65 MeV Electron linear Accelerator Using an Ionization Ion-Electron Chamber.

Nucl. Instrum. Methods Phys. Res. Sect. A 270, 542-549 (1988)
DOI
Open Access Green as soon as Postprint is submitted to ZB.
Ionometric measurements in the beams of linear accelerators can be carried out even at high pulse dose rates if one considers the ionic recombination within the ionization chamber. It is shown that this in turn provides the knowledge of the electron component collected in the chamber. We used a chamber developed for that very purpose, which allows a direct measurement of the electron component. Results up to a pulse dose of 1.1 × 10−2 C kg−1 (approximately 43 R) are presented.
Altmetric
Additional Metrics?
Edit extra informations Login
Publication type Article: Journal article
Document type Scientific Article
Corresponding Author
ISSN (print) / ISBN 0168-9002
e-ISSN 1872-9576
Quellenangaben Volume: 270, Issue: 2-3, Pages: 542-549 Article Number: , Supplement: ,
Publisher Elsevier
Non-patent literature Publications
Reviewing status Peer reviewed