Gómez-Ros, J.M.* ; de Carlan, L.* ; Franck, D.* ; Gualdrini, G.* ; Lis, M.* ; López, M.A.* ; Moraleda, M.* ; Zankl, M. ; Badal, A.* ; Capello, K.* ; Cowan, P.* ; Ferrari, P.* ; Heide, B.* ; Henniger, J.* ; Hooley, V.* ; Hunt, J.* ; Kinase, S.* ; Kramer, G.H.* ; Löhnert, D.* ; Lucas, S.* ; Nuttens, V.* ; Packer, L.W.* ; Reichelt, U.* ; Vrba, T.* ; Sempau, J.* ; Zhang, B.*

Monte Carlo modelling of Germanium detectors for the measurement of low energy photons in internal dosimetry: Results of an international comparison.

Radiat. Meas. 43, 510-515 (2008)
DOI
Open Access Green as soon as Postprint is submitted to ZB.
This communication summarizes the results concerning the Monte Carlo (MC) modelling of Germanium detectors for the measurement of low energy photons arising from the “International comparison on MC modelling for in vivo measurement of Americium in a knee phantom” organized within the EU Coordination Action CONRAD (Coordinated Network for Radiation Dosimetry) as a joint initiative of EURADOS working groups 6 (computational dosimetry) and 7 (internal dosimetry). MC simulations proved to be an applicable way to obtain the calibration factor that needs to be used for in vivo measurements.
Altmetric
Additional Metrics?
Edit extra informations Login
Publication type Article: Journal article
Document type Scientific Article
Corresponding Author
Keywords Ge detectors; Monte Carlo simulations; Voxel phantoms; Internal dosimetry
ISSN (print) / ISBN 1350-4487
e-ISSN 1879-0925
Quellenangaben Volume: 43, Issue: 2-6, Pages: 510-515 Article Number: , Supplement: ,
Publisher Elsevier
Non-patent literature Publications