Raster-scan optoacoustic mesoscopy in the 25-125 MHz range.

Opt. Lett. 38, 2472-2474 (2013)
DOI
Open Access Green as soon as Postprint is submitted to ZB.
We developed a raster-scan acoustic resolution broadband optoacoustic mesoscopy system and investigated the imaging performance using ultrasonic frequencies up to 125 MHz. The developed system achieves 7 μm axial resolution and transverse resolution of 30 μm reaching depths of at least 5 mm. This unprecedented performance is achieved by operating at out-of-focus ultrasonic detection and tomographic reconstruction. We demonstrate the limits reached due to the width of the laser pulse employed and showcase the technique on drosophila fly and drosophila pupae ex vivo.
Altmetric
Additional Metrics?
Edit extra informations Login
Publication type Article: Journal article
Document type Scientific Article
Corresponding Author
Keywords In-vivo ; Photoacoustic Microscopy ; High-resolution ; Tomography ; Deep
ISSN (print) / ISBN 0146-9592
e-ISSN 1539-4794
Journal Optics Letters
Quellenangaben Volume: 38, Issue: 14, Pages: 2472-2474 Article Number: , Supplement: ,
Publisher Optical Society of America (OSA)
Non-patent literature Publications
Reviewing status Peer reviewed