Open Access Green as soon as Postprint is submitted to ZB.
Intercomparison of radiation protection instrumentation in a pulsed neutron field.
Nucl. Instrum. Methods Phys. Res. Sect. A 737, 203-213 (2014)
In the framework of the EURADOS working group 11, an intercomparison of active neutron survey meters was performed in a pulsed neutron field (PNF). The aim of the exercise was to evaluate the performances of various neutron instruments, including commercially available rem-counters, personal dosemeters and instrument prototypes. The measurements took place at the cyclotron of the Helmholtz-Zentrum Berlin für Materialien und Energie GmbH. The cyclotron is routinely used for proton therapy of ocular tumours, but an experimental area is also available. For the therapy the machine accelerates protons to 68 MeV. The interaction of the proton beam with a thick tungsten target produces a neutron field with energy up to about 60 MeV. One interesting feature of the cyclotron is that the beam can be delivered in bursts, with the possibility to modify in a simple and flexible way the burst length and the ion current. Through this possibility one can obtain radiation bursts of variable duration and intensity. All instruments were placed in a reference position and irradiated with neutrons delivered in bursts of different intensity. The analysis of the instrument response as a function of the burst charge (the total electric charge of the protons in the burst shot onto the tungsten target) permitted to assess for each device the dose underestimation due to the time structure of the radiation field. The personal neutron dosemeters were exposed on a standard PMMA slab phantom and the response linearity was evaluated.
Impact Factor
Scopus SNIP
Web of Science
Times Cited
Times Cited
Scopus
Cited By
Cited By
Altmetric
1.316
1.419
29
30
Annotations
Special Publikation
Hide on homepage
Publication type
Article: Journal article
Document type
Scientific Article
Keywords
Neutron detection; Neutron survey meter; Proton beam; Pulsed neutron field; Rem Counter Linus; Recombination Chamber; Range; Meter; Sensitivity; Calibration; Dosimeters
Language
english
Publication Year
2014
HGF-reported in Year
2014
ISSN (print) / ISBN
0168-9002
e-ISSN
1872-9576
Quellenangaben
Volume: 737,
Pages: 203-213
Publisher
Elsevier
Reviewing status
Peer reviewed
Institute(s)
Institute of Radiation Protection (ISS)
POF-Topic(s)
30504 - Mechanisms of Genetic and Environmental Influences on Health and Disease
Research field(s)
Radiation Sciences
PSP Element(s)
G-501100-001
WOS ID
WOS:000329404500030
Scopus ID
84890454597
Erfassungsdatum
2014-01-30