Open Access Green as soon as Postprint is submitted to ZB.
Study of the neutron response of a G-M counter.
Radiat. Prot. Dosim. 13, 373-375 (1985)
The use of a commercial G-M counter as a neutron insensitive device in mixed neutron/photon fields was explored experimentally. The counter was operated in the proportional region, and the event size spectra were measured and analysed. Pulse height spectra were collected over a dynamic range of 104. Considerable experimental difficulties were encountered because of the steep slope of the relationship between pulse height and collection voltage in the proportional region. Values of k(u) for monoenergetic neutrons of 0.57 and 1.5 MeV were obtained from an analysis of the pulse height spectra. They are compared with data available from other techniques. Theoretical considerations are given to explain qualitatively the spectra caused by the nuclear reactions of the neutrons in the counter gas and wall.
Impact Factor
Scopus SNIP
Altmetric
0.000
0.000
Annotations
Special Publikation
Hide on homepage
Publication type
Article: Journal article
Document type
Scientific Article
Language
english
Publication Year
1985
HGF-reported in Year
0
ISSN (print) / ISBN
0144-8420
e-ISSN
1742-3406
Journal
Radiation Protection Dosimetry
Quellenangaben
Volume: 13,
Issue: 1-4,
Pages: 373-375
Publisher
Oxford University Press
Publishing Place
Oxford
Reviewing status
Peer reviewed
Institute(s)
Institute of Radiation Protection (ISS)
Scopus ID
0022365977
Erfassungsdatum
1985-12-31