Iodine determination in biological materials by ICP-MS.

Mikrochimica acta 116, 205-209 (1994)
DOI
Open Access Green as soon as Postprint is submitted to ZB.
The present work describes an analytical procedure for the determination of the total iodine content in biological materials (serum, milk, plants, tissues etc.). Liquid samples can be directly analyzed after dilution, if necessary, by ICP-MS. Milk powder, plants and tissues were dissolved by using a modified simple Schöninger combustion, subsequently the residue was taken up in 0.1 mol/l NaOH and this solution was analyzed by ICP-MS. The detection limit is in the range of 0.01 μg/l. The method was tested using different CRMs certified for the iodine content.
Altmetric
Additional Metrics?
Edit extra informations Login
Publication type Article: Journal article
Document type Scientific Article
Keywords Biological Material ; Crms ; Icp-ms ; Iodine ; Schöninger Combustion
e-ISSN 0026-3672
Quellenangaben Volume: 116, Issue: 4, Pages: 205-209 Article Number: , Supplement: ,
Publisher Springer
Reviewing status Peer reviewed