Diagnostic X ray spectra behind phantom and antiscatter grid.

Radiat. Prot. Dosim. 43, 151-154 (1992)
DOI
Open Access Green as soon as Postprint is submitted to ZB.
Spectra of the total direct and scattered radiation behind a phantom (30 x 30 x 20 cm3) and an antiscatter grid (8/40) were measured. The results show that the antiscatter grid has a remarkable impact on the spectral distribution and that spectra used up to now for standard sensitometry of film-screen systems differ widely from the spectra as they are in practice; the latter can be simulated by filtration.