PuSH - Publication Server of Helmholtz Zentrum München

On-line preconcentration of arsenic from water samples with an anion-exchanger column coupled to a hydride generation ICP system.

Fresen. J. Anal. Chem. 343, 373-377 (1992)
DOI
Open Access Green as soon as Postprint is submitted to ZB.
An on-line anion-exchange preconcentration hydride generation ICP system for the determination of total inorganic arsenic in water is described. The column was packed with strongly basic anion-exchange resin (AG 1-X8). Experimental conditions including pH of the sample solution, eluent, flow rate of eluent, oxidation states of arsenic and competing anion ions were studied. Compared with the conventional continuous hydride generation ICP, a 9.2-fold improvement in sensitivity was obtained with RSD 1-2% at 100 ng/ml. The detection limit (3σ) was 0.08 ng/ml. The recoveries in water samples were satisfactory. The system provides complete automation of sample loading, eluting and regenerating of the resin.
Altmetric
Additional Metrics?
Edit extra informations Login
Publication type Article: Journal article
Document type Scientific Article
Corresponding Author
ISSN (print) / ISBN 0937-0633
e-ISSN 1432-1130
Quellenangaben Volume: 343, Issue: 4, Pages: 373-377 Article Number: , Supplement: ,
Publisher Springer
Non-patent literature Publications
Reviewing status Peer reviewed
Institute(s) Institute of Ecological Chemistry (IOEC)