PuSH - Publication Server of Helmholtz Zentrum München

A 400 rad flash x ray system for pulse radiolysis.

Rev. Sci. Instruments 45, 650-652 (1974)
Publ. Version/Full Text DOI
Closed
Open Access Green as soon as Postprint is submitted to ZB.
The design and working principle of a flash x ray system are described. A two stage Marx generator is used, the charging voltage is 150 kV and the stored energy 1125 J. Anode and cathode are arranged coaxially. The anode consists of a tungsten coated aluminium tube of 20 mm i.d. into which the substances to be irradiated are placed. Measurements of electrical and radiation parameters are described. The x ray dose per pulse is 400 rad, the half width of the pulse is 0.2 μsec and the mean quantum energy is 44 keV. An example for the application of the system in microsecond pulse radiolysis is presented.
Impact Factor
Scopus SNIP
Altmetric
0.000
0.000
Tags
Annotations
Special Publikation
Hide on homepage

Edit extra information
Edit own tags
Private
Edit own annotation
Private
Hide on publication lists
on hompage
Mark as special
publikation
Publication type Article: Journal article
Document type Scientific Article
Language english
Publication Year 1974
HGF-reported in Year 0
ISSN (print) / ISBN 0034-6748
e-ISSN 1089-7623
Quellenangaben Volume: 45, Issue: 5, Pages: 650-652 Article Number: , Supplement: ,
Publisher American Institute of Physics (AIP)
Reviewing status Peer reviewed
Institute(s) Institut für Strahlenbotanik
Scopus ID 0016248342
Erfassungsdatum 1974-11-27