Open Access Green as soon as Postprint is submitted to ZB.
Ein Vergleich der Anzeige von Filmdosimetern mit der von gleichzeitig getragenen Stabdosimetern.
A composition of the readings of film- and pocket dosimetry.
Radiologe 19, 454-456 (1979)
An inquiry to compare the readings of film- and pocket dosimeters worn by the same individual at the same time showed similar results, a fact which was found to be very satisfying. Of the 2132 dosimeters compared, only 46 (2.2%) were outside the standard error of -30/+50% permitted for film dosimetry. The majority of these (37) can be explained to be due to the different type of energy dependence of these 2 dosimeter types. The total dosis measured reflected an even better conformity, differing only by 11.3% or 14.4% depending on the mode of calculation, the difference being partially due to the better sensitivity of the film dosimeters to low energy beta radiation. An additional result of the study is the conclusion to recommend more careful use and recording of pocket dosimetry.
Altmetric
Additional Metrics?
Edit extra informations
Login
Publication type
Article: Journal article
Document type
Scientific Article
ISSN (print) / ISBN
0033-832X
e-ISSN
1432-2102
Journal
Radiologe, Der
Quellenangaben
Volume: 19,
Issue: 10,
Pages: 454-456
Publisher
Springer
Non-patent literature
Publications
Reviewing status
Peer reviewed
Institute(s)
Institute of Radiation Protection (ISS)