PuSH - Publication Server of Helmholtz Zentrum München

Bergmeier, F. ; Volnhals, M.* ; Wielunski, M. ; Rühm, W.

Simulation and calibration of an active neutron dosemeter.

Radiat. Prot. Dosim. 161, 126-129 (2014)
DOI
Open Access Green as soon as Postprint is submitted to ZB.
Here the latest development stages of the HMGU active neutron dosemeter are presented. This work includes the comparison of the dosemeter's response function, calculated with Geant4, and the measurements in monoenergetic neutron fields at the Physikalisch Technische Bundesanstalt in Braunschweig, Germany. These results were used to match the response function and the count-to-dose conversion factors of the dosemeter to the H-p(10) personal dose equivalent.
Impact Factor
Scopus SNIP
Web of Science
Times Cited
Scopus
Cited By
Altmetric
0.861
0.802
6
5
Tags
Annotations
Special Publikation
Hide on homepage

Edit extra information
Edit own tags
Private
Edit own annotation
Private
Hide on publication lists
on hompage
Mark as special
publikation
Publication type Article: Journal article
Document type Scientific Article
Language english
Publication Year 2014
HGF-reported in Year 2014
ISSN (print) / ISBN 0144-8420
e-ISSN 1742-3406
Quellenangaben Volume: 161, Issue: 1-4, Pages: 126-129 Article Number: , Supplement: ,
Publisher Oxford University Press
Publishing Place Oxford
Reviewing status Peer reviewed
POF-Topic(s) 30504 - Mechanisms of Genetic and Environmental Influences on Health and Disease
Research field(s) Radiation Sciences
PSP Element(s) G-501100-001
Scopus ID 84920702252
Erfassungsdatum 2014-11-17