PuSH - Publication Server of Helmholtz Zentrum München

Bergmeier, F. ; Volnhals, M.* ; Wielunski, M. ; Rühm, W.

Simulation and calibration of an active neutron dosemeter.

Radiat. Prot. Dosim. 161, 126-129 (2014)
DOI
Open Access Green as soon as Postprint is submitted to ZB.
Here the latest development stages of the HMGU active neutron dosemeter are presented. This work includes the comparison of the dosemeter's response function, calculated with Geant4, and the measurements in monoenergetic neutron fields at the Physikalisch Technische Bundesanstalt in Braunschweig, Germany. These results were used to match the response function and the count-to-dose conversion factors of the dosemeter to the H-p(10) personal dose equivalent.
Altmetric
Additional Metrics?
Edit extra informations Login
Publication type Article: Journal article
Document type Scientific Article
Corresponding Author
ISSN (print) / ISBN 0144-8420
e-ISSN 1742-3406
Quellenangaben Volume: 161, Issue: 1-4, Pages: 126-129 Article Number: , Supplement: ,
Publisher Oxford University Press
Publishing Place Oxford
Non-patent literature Publications
Reviewing status Peer reviewed