Simulation and calibration of an active neutron dosemeter.
Radiat. Prot. Dosim. 161, 126-129 (2014)
Here the latest development stages of the HMGU active neutron dosemeter are presented. This work includes the comparison of the dosemeter's response function, calculated with Geant4, and the measurements in monoenergetic neutron fields at the Physikalisch Technische Bundesanstalt in Braunschweig, Germany. These results were used to match the response function and the count-to-dose conversion factors of the dosemeter to the H-p(10) personal dose equivalent.
Altmetric
Additional Metrics?
Edit extra informations
Login
Publication type
Article: Journal article
Document type
Scientific Article
ISSN (print) / ISBN
0144-8420
e-ISSN
1742-3406
Journal
Radiation Protection Dosimetry
Quellenangaben
Volume: 161,
Issue: 1-4,
Pages: 126-129
Publisher
Oxford University Press
Publishing Place
Oxford
Reviewing status
Peer reviewed
Institute(s)
Institute of Radiation Protection (ISS)