PuSH - Publication Server of Helmholtz Zentrum München

Wolf, M.* ; Riedlinger, I.* ; Lehmann, R. ; Häring, H.-U. ; Schleicher, E. ; Peter, A.

Comparison of the automated KRYPTOR chromogranin A assay with the DAKO ELISA.

Clin. Lab. 60, 2103-2106 (2014)
DOI PMC
Open Access Green as soon as Postprint is submitted to ZB.
Background: Currently available assays for the widely used marker for neuroendocrine tumors chromogranin A have a weak comparability, involve manual work, and require batch processing of the samples. In this study, we evaluated the automated chromogranin A KRYPTOR assay compared with the widely used DAKO ELISA. Methods: 83 samples were measured with the DAKO ELISA (EDTA plasma) and the KRYPTOR assay (serum), since different sample materials are recommended. Furthermore, different sample materials were compared. Results: The results between the two assays were highly correlated KRYPTOR serum (ng/mL) = 2.887 ELISA EDTA (U/L) + 5.028; r = 0.99). The inter-assay variation for the KRYPTOR assay was determined as 5.56% at low (95 ng/mL; n = 34) and 6.21% at high (530 ng/mL; n = 33) chromogranin A concentrations. EDTA plasma samples revealed significantly lower results than serum in the KRYPTOR assays and could not be used. Conclusions: In conclusion, the new chromogranin A KRYPTOR assay is fast, reliable, and compares well to an established test after adaptation of sample material and reported units.
Impact Factor
Scopus SNIP
Web of Science
Times Cited
Scopus
Cited By
Altmetric
1.084
0.480
4
3
Tags
Annotations
Special Publikation
Hide on homepage

Edit extra information
Edit own tags
Private
Edit own annotation
Private
Hide on publication lists
on hompage
Mark as special
publikation
Publication type Article: Journal article
Document type Scientific Article
Keywords Chromogranin A ; Assay ; Kryptor ; Tumor Marker; Tumors
Language english
Publication Year 2014
HGF-reported in Year 2015
ISSN (print) / ISBN 1433-6510
Quellenangaben Volume: 60, Issue: 12, Pages: 2103-2106 Article Number: , Supplement: ,
Publisher Clinical Laboratory Publ.
Publishing Place Mainz
Reviewing status Peer reviewed
POF-Topic(s) 90000 - German Center for Diabetes Research
Research field(s) Helmholtz Diabetes Center
PSP Element(s) G-502400-001
PubMed ID 25651748
Erfassungsdatum 2015-02-16