Optoacoustic characterization of broadband directivity patterns of capacitive micromachined ultrasonic transducers.
J. Biomed. Opt. 22:041005 (2017)
Frequency characteristics of ultrasound detectors used in optoacoustic tomography have a major impact on imaging performance. It is common practice to select transducers based on their sensitivity at the central frequency and under normal incidence. However, the bandwidth and angular sensitivity play an equally important role in establishing the quality and accuracy of the reconstructed images. Here, we developed a calibrated optoacoustic characterization method specifically tailored for broadband measurements of the angular transducer sensitivity (directivity). Ultrawideband omnidirectional optoacoustic responses were generated by uniformly illuminating thin absorbing sutures with nanosecond laser pulses and characterized with a needle hydrophone. This calibrated optoacoustic source was used to characterize the frequency dependence of the angular response by a conventional piezoelectric transducer (PZT) and a capacitive micromachined ultrasonic transducer (cMUT) with similar size and central frequency. Furthermore, both transducers had no preamplification electronics directly attached to the detection elements. While the PZT presented a 7.8 dB sensitivity advantage at normal incidence, it was able to provide detectable signal-to-noise levels only at incidence angles of up to 20 deg whereas the cMUT maintained reasonable sensitivity levels and broadband response at incidence angles of 40 deg and beyond. We further experimentally showcase a reduction in the limited-view image artifacts resulting from the broader acceptance angle of the cMUT.
Impact Factor
Scopus SNIP
Web of Science
Times Cited
Scopus
Cited By
Altmetric
Publication type
Article: Journal article
Document type
Scientific Article
Thesis type
Editors
Keywords
directivity; angular sensitivity; optoacoustic characterization; capacitive micromachined ultrasonic transducer; conventional
piezoelectric transducer; Calibration Technique; Cmut Array; Tomography; Sensitivity; Probe; Hydrophones; Standard; System
Keywords plus
Language
Publication Year
2017
Prepublished in Year
2016
HGF-reported in Year
2016
ISSN (print) / ISBN
1083-3668
e-ISSN
1560-2281
ISBN
Book Volume Title
Conference Title
Conference Date
Conference Location
Proceedings Title
Quellenangaben
Volume: 22,
Issue: 4,
Pages: ,
Article Number: 041005
Supplement: ,
Series
Publisher
SPIE
Publishing Place
Bellingham
Day of Oral Examination
0000-00-00
Advisor
Referee
Examiner
Topic
University
University place
Faculty
Publication date
0000-00-00
Application date
0000-00-00
Patent owner
Further owners
Application country
Patent priority
Reviewing status
Peer reviewed
POF-Topic(s)
30205 - Bioengineering and Digital Health
Research field(s)
Enabling and Novel Technologies
PSP Element(s)
G-505590-001
Grants
Copyright
Erfassungsdatum
2016-11-17