Tischenko, O. ; Saeednejad, N.* ; Hoeschen, C.*

Novel method to calibrate CT scanners with a conic probe body.

Proc. SPIE 10132:101322Q (2017)
Postprint DOI
Open Access Green
A new and simple object for calibrating tomographic scanners has been proposed. Instead of a conventional high-density ball as an object for calibration, we propose a high-density conic body. The cone is advantageous compare to the ball both because of its easy availability (uncomplicated manufacturing) and the straightforward and less error-prone analysis necessary for the identification of a space point (ball’s center vs. cone apex). Applying the conic body instead of a ball as a calibration object enables to reduce calibration errors substantially. Additionally we propose an efficient way to determine the discrepancy between ideal and misaligned positions of the detector that may be crucial for the quality of the reconstruction.
Impact Factor
Scopus SNIP
Altmetric
0.000
0.384
Tags
Annotations
Special Publikation
Hide on homepage

Edit extra information
Edit own tags
Private
Edit own annotation
Private
Hide on publication lists
on hompage
Mark as special
publikation
Publication type Article: Journal article
Document type Scientific Article
Keywords x-ray tomography, cone beam CT, calibration
Language english
Publication Year 2017
HGF-reported in Year 2017
ISSN (print) / ISBN 0277-786X
e-ISSN 1996-756X
Conference Title Medical Imaging 2017: Physics of Medical Imaging
Conference Date 11 February 2017
Conference Location Orlando, Florida, United States
Quellenangaben Volume: 10132, Issue: , Pages: , Article Number: 101322Q Supplement: ,
Publisher SPIE
Reviewing status Peer reviewed
POF-Topic(s) 30203 - Molecular Targets and Therapies
30504 - Mechanisms of Genetic and Environmental Influences on Health and Disease
Research field(s) Radiation Sciences
PSP Element(s) G-501391-001
G-501100-008
Scopus ID 85020413566
Erfassungsdatum 2017-03-20