PuSH - Publication Server of Helmholtz Zentrum München

Mieth, M.* ; Schubert, J.K.* ; Gröger, T.M. ; Sabel, B.* ; Kischkel, S.* ; Fuchs, P.* ; Hein, D.* ; Zimmermann, R. ; Miekisch, W.*

Automated needle trap heart-cut GC/MS and needle trap comprehensive two-dimensional GC/TOF-MS for breath gas analysis in the clinical environment.

Anal. Chem. 82, 2541-2551 (2010)
DOI PMC
Open Access Green as soon as Postprint is submitted to ZB.
This study was intended to evaluate low-volume (20 mL) multibed needle trap (NTD) sampling combined with heart-cut gas chromatography/mass spectrometry (GC/MS) and comprehensive two-dimensional gas chromatography/time-of-flight mass spectrometry (GC x GC/TOF-MS) for trace gas analysis under clinical conditions. NTDs, high-throughput automatic desorption and separation systems, were tested in vitro and within a study in 11 patients undergoing cardiac surgery with respect to reproducibility, reliability, and clinical applicability. NTD-heart-cut GC/MS analysis of standard mixtures containing different volatile organic compounds (VOCs) yielded relative standard deviations (RSDs) from 4.0% to 18.5%. Substance adsorption was stable for 1 day if NTDs were closed on both ends and was stable for approximately 7.8 h when NTD tip ends had to be left open during autosampler storage. Even in the presence of high concentrations of contaminants linearity of heart-cut GC/MS was conserved. In patients' breath potential biomarkers could be determined even in the presence of very high concentrations of sevoflurane. Profiles of blood-borne biomarkers, intravenous drugs, and clinical contaminants were characterized. Comprehensive GC x GC/TOF-MS may be used as a screening tool for new biomarkers, if patterns are generated from deconvoluted normalized areas. Needle trap sampling in combination with hyphenated chromatographic techniques can thus be used to provide well-tailored solutions for complex problems occurring in clinical breath analysis.
Impact Factor
Scopus SNIP
Web of Science
Times Cited
Scopus
Cited By
Altmetric
5.214
2.320
109
115
Tags
Annotations
Special Publikation
Hide on homepage

Edit extra information
Edit own tags
Private
Edit own annotation
Private
Hide on publication lists
on hompage
Mark as special
publikation
Publication type Article: Journal article
Document type Scientific Article
Keywords Flight mass-spectrometry; X GC-tofms; Volatile organic-compounds; Sorption trap; Chromatography; Biomarkers; Devices; Samples; Flow; Chemiluminescence
Language english
Publication Year 2010
HGF-reported in Year 2010
ISSN (print) / ISBN 0003-2700
e-ISSN 1520-6882
Quellenangaben Volume: 82, Issue: 6, Pages: 2541-2551 Article Number: , Supplement: ,
Publisher American Chemical Society (ACS)
Reviewing status Peer reviewed
Institute(s) Institute of Ecological Chemistry (IOEC)
Cooperation Group Comprehensive Molecular Analytics (CMA)
POF-Topic(s)
Research field(s)
PSP Element(s) G-505100-002
PubMed ID 20170082
Scopus ID 77949866490
Erfassungsdatum 2010-03-15