PuSH - Publication Server of Helmholtz Zentrum München

Bandalo, V. ; Greiter, M. ; Brönner, J. ; Hödlmoser, H.*

ISO 4037-2019 validation of radiation qualities by means of half-value layer and. Hp(10) dosimetry.

Radiat. Prot. Dosim. 187, 438-450 (2019)
DOI PMC
Open Access Green as soon as Postprint is submitted to ZB.
Radiation qualities were characterised and validated at the fully automated X-ray calibration facility of the Individual Monitoring Service at Helmholtz Zentrum Munchen by using half-value layer and H-p(10) dosimetry approaches specified in the updated ISO 4037:2019 standard. As the ISO 4037 contains a somewhat vague description of the half-value layer procedure, we extended it to be more constrained and thus less subjective in its implementation. We specify both the measurement and data analysis steps performed in order to provide reproducible half-value layer results and compare the results with the H-p(10) dosimetry-based validation approach specified in the ISO 4037 as an alternative for field validation. Finally, we discuss the implications of our results and the extended procedure on validation and the recent changes to the ISO 4037.
Impact Factor
Scopus SNIP
Web of Science
Times Cited
Scopus
Cited By
Altmetric
0.831
0.687
7
6
Tags
Annotations
Special Publikation
Hide on homepage

Edit extra information
Edit own tags
Private
Edit own annotation
Private
Hide on publication lists
on hompage
Mark as special
publikation
Publication type Article: Journal article
Document type Scientific Article
Language english
Publication Year 2019
HGF-reported in Year 2019
ISSN (print) / ISBN 0144-8420
e-ISSN 1742-3406
Quellenangaben Volume: 187, Issue: 4, Pages: 438-450 Article Number: , Supplement: ,
Publisher Oxford University Press
Publishing Place Oxford
Reviewing status Peer reviewed
POF-Topic(s) 30504 - Mechanisms of Genetic and Environmental Influences on Health and Disease
Research field(s) Radiation Sciences
PSP Element(s) A-660200-001
PubMed ID 31650180
Erfassungsdatum 2020-01-30