A quantitative evaluation of deformable image registration based on MV cone beam CT images: Impact of deformation magnitudes and image modalities.
Phys. Med. 71, 82-87 (2020)
Background and purpose: To evaluate the impact of deformation magnitude and image modality on deformable-image-registration (DIR) accuracy using Halcyon megavoltage cone beam CT images (MVCBCT).Materials and methods: Planning CT images of an anthropomorphic Head phantom were aligned rigidly with MVCBCT and re-sampled to achieve the same resolution, denoted as pCT. MVCBCT was warped with twenty simulated pre-known virtual deformation fields (T-i, i = 1-20) with increasing deformation magnitudes, yielding warped CBCT (wCBCT). The pCT and MVCBCT were registered to wCBCT respectively (Multi-modality and Unimodality DIR), generating deformation vector fields V-i and V-i' (i = 1-20). V-i and V-i' were compared with Ti respectively to assess the DIR accuracy geometrically. In addition, V-i, T-i, and V-i' were applied to pCT, generating deformed CT (dCT(i)), ground-truth CT (G(i)) and deformed CT' (dCT(i)') respectively. The Hounsfield Unit (HU) on these virtual CT images were also compared.Results: The mean errors of vector displacement increased with the deformation magnitude. For deformation magnitudes between 2.82 mm and 7.71 mm, the errors of uni-modality DIR were 1.16 mm similar to 1.73 mm smaller than that of multi-modality (p = 0.0001, Wilcoxon signed rank test). DIR could reduce the maximum signed and absolute HU deviations from 70.8 HU to 11.4 HU and 208 HU to 46.2 HU respectively.Conclusions: As deformation magnitude increases, DIR accuracy continues to deteriorate and uni-modality DIR consistently outperformed multi-modality DIR. DIR-based adaptive radiotherapy utilizing the noisy MVCBCT images is only conditionally applicable with caution.
Altmetric
Additional Metrics?
Edit extra informations
Login
Publication type
Article: Journal article
Document type
Scientific Article
Keywords
Mvcbct ; Adaptive Radiotherapy ; Deformable Image Registration; Computed-tomography; Radiotherapy; Validation; Algorithms; Accuracy; Abdomen
ISSN (print) / ISBN
1120-1797
e-ISSN
1724-191X
Quellenangaben
Volume: 71,
Pages: 82-87
Publisher
Elsevier
Publishing Place
The Boulevard, Langford Lane, Kidlington, Oxford Ox5 1gb, Oxon, England
Non-patent literature
Publications
Reviewing status
Peer reviewed
Institute(s)
Institute of Radiation Medicine (IRM)