Corrected equation for the concentrations in the drift tube of a proton transfer reaction-mass spectrometer (PTR-MS).

Int. J. Mass Spectrom. 264, 92-95 (2007)
Postprint DOI
Open Access Green
A corrected equation for the concentrations in the drift tube of a proton transfer reaction-mass spectrometer (PTR-MS) is presented which accounts for the different mobility of primary ions and protonated analyte previous termvolatilenext term organic compounds (VOCs). As a consequence, the calculation of VOC concentrations from the measured count rates requires as well a correction factor, equalling the ratio of (i) the mobility of the protonated analyte VOCs to (ii) the mobility of the gases used to determine the transmission efficiency. However, such data do essentially not exist for the relevant VOCs. Published mobility data for small inorganic ions suggest that the correction can be larger than 20% and hence be a significant contribution to the overall accuracy of calculated concentrations. The correction emphasises the need to calibrate the PTR-MS in order to determine VOC concentrations accurately.
Impact Factor
Scopus SNIP
Web of Science
Times Cited
Scopus
Cited By
Altmetric
2.337
0.921
14
17
Tags
Annotations
Special Publikation
Hide on homepage

Edit extra information
Edit own tags
Private
Edit own annotation
Private
Hide on publication lists
on hompage
Mark as special
publikation
Publication type Article: Journal article
Document type Scientific Article
Keywords Proton transfer reaction-mass spectrometry; previous termVolatile organic compounds
Language
Publication Year 2007
HGF-reported in Year 2007
ISSN (print) / ISBN 1387-3806
e-ISSN 1873-2798
Quellenangaben Volume: 264, Issue: 1, Pages: 92-95 Article Number: , Supplement: ,
Publisher Elsevier
Reviewing status Peer reviewed
PSP Element(s) G-501100-007
Scopus ID 34247860745
Erfassungsdatum 2007-06-15