Open Access Green as soon as Postprint is submitted to ZB.
Application of the Steam Jet Aerosol Collector (SJAC) for differential measurements of organic compounds in the gas vs. particle phase.
Int. J. Environ. Pollut. 39, 223-232 (2009)
A modified Steam Jet Aerosol Collector (SJAC) for semivolatile organic compounds has been investigated. Gas phase and particle phase were investigated separately. The SJAC is based on a water condensational growth sampling system. For determination of gaseous compounds an XAD adsorbent cartridge was applied after particle collection. Collection particle efficiency was examined using hydrophobic soot and octacosane particles. To estimate the bias caused by the transfer of components from the gaseous to the particle fraction, individual gaseous organic with different polarity were examined. High collection efficiency and negligible cross-contamination of gaseous compounds in the particle fraction was observed.
Altmetric
Additional Metrics?
Edit extra informations
Login
Publication type
Article: Journal article
Document type
Scientific Article
Keywords
Semi volatile organic compounds; Aerosol sampling; Gas/particle portioning
ISSN (print) / ISBN
0957-4352
e-ISSN
1741-5101
Quellenangaben
Volume: 39,
Issue: 3-4,
Pages: 223-232
Publisher
Inderscience Enterprises Limited
Non-patent literature
Publications
Reviewing status
Peer reviewed
Institute(s)
Institute of Ecological Chemistry (IOEC)
Institute of Lung Biology (LHI)
Cooperation Group Comprehensive Molecular Analytics (CMA)
Institute of Lung Biology (LHI)
Cooperation Group Comprehensive Molecular Analytics (CMA)