PuSH - Publication Server of Helmholtz Zentrum München

Zorloni, G.* ; Bosmans, G.* ; Brall, T. ; Caresana, M.* ; De Saint-Hubert, M.* ; Domingo, C.* ; Ferrante, C.* ; Ferrulli, F.* ; Kopec, R.* ; Leidner, J.* ; Mares, V. ; Nabha, R.* ; Olko, P.* ; Caballero-Pacheco, M.A.* ; Rühm, W. ; Silari, M.* ; Stolarczyk, L.* ; Swakon, J.* ; Tisi, M. ; Trinkl, S.* ; van Hoey, O.* ; Vilches-Freixas, G.*

Eurados rem-counter intercomparison at Maastro proton therapy centre: Comparison with literature data.

Radiat. Prot. Dosim. 198, 1471-1475 (2022)
DOI PMC
Open Access Green as soon as Postprint is submitted to ZB.
The Maastro Proton Therapy Centre is the first European facility housing the Mevion S250i Hyperscan synchrocyclotron. The proximity of the accelerator to the patient, the presence of an active pencil beam delivery system downstream of a passive energy degrader and the pulsed structure of the beam make the Mevion stray neutron field unique amongst proton therapy facilities. This paper reviews the results of a rem-counter intercomparison experiment promoted by the European Radiation Dosimetry Group at Maastro and compares them with those at other proton therapy facilities. The Maastro neutron H*(10) in the room (100-200 μSv/Gy at about 2 m from the isocentre) is in line with accelerators using purely passive or wobbling beam delivery modalities, even though Maastro shows a dose gradient peaked near the accelerator. Unlike synchrotron- and cyclotron-based facilities, the pulsed beam at Maastro requires the employment of rem-counters specifically designed to withstand pulsed neutron fields.
Altmetric
Additional Metrics?
Edit extra informations Login
Publication type Article: Journal article
Document type Scientific Article
Corresponding Author
ISSN (print) / ISBN 0144-8420
e-ISSN 1742-3406
Quellenangaben Volume: 198, Issue: 19, Pages: 1471-1475 Article Number: , Supplement: ,
Publisher Oxford University Press
Publishing Place Oxford
Non-patent literature Publications
Reviewing status Peer reviewed