Suarez-Suarez, M. ; Maya-Manzano, J.M. ; Clot, B.* ; Graber, M.J.* ; Sallin, C.* ; Tummon, F.* ; Buters, J.T.M.

Accuracy of a hand-held resistance-free flowmeters for flow adjustments of Hirst-Type pollen traps.

Aerobiologia 39, 143–148 (2023)
Publ. Version/Full Text DOI
Open Access Hybrid
Creative Commons Lizenzvertrag
The Hirst-type pollen trap (1952) is the most common device for aerobiological measurements of pollen and fungal spores in ambient air. In the 1960s the in-line flowmeter was removed and studies since then showed considerable variability in the airflow pumped through the instrument when using hand-held rotameters with an internal airflow resistance to adjust flowrates. To avoid this problem, our study compared the variability of airflow rates of Hirst-type traps when using commercially-available low-resistance airflow meters (heat anemometers) at various timescales. Experiments were conducted in Munich (Germany) and Payerne (Switzerland), using 4 different easyFlux® instruments and 6 Hirst-type pollen traps. Measurements were taken on an hourly basis from dawn to dusk at both locations, and in addition at Payerne, weekly observations over a period of one year. When using the common hand-held rotameters (with airflow resistance) the flow was 28.3% lower than with resistance-free flowmeter (i.e., measured 10 L/min which was in reality 12.8 L/min). The coefficient of variation between the four easyFlux® devices ranged from 0.32% to 1.55% over one day and from 2.88% to 8.17% over an entire year. Some of the traps showed surprising flow variations during the day. Furthermore, flowrates deviated more when measurements were made at the point where the double-sided tape is behind the orifice than elsewhere on the drum. The measurements away from this point are representative of the flow rates for most of the period of operation and flow calibration should thus be carried out away from this point, contrary to the current procedure.
Impact Factor
Scopus SNIP
Web of Science
Times Cited
Altmetric
2.000
0.778
1
Tags
Annotations
Special Publikation
Hide on homepage

Edit extra information
Edit own tags
Private
Edit own annotation
Private
Hide on publication lists
on hompage
Mark as special
publikation
Publication type Article: Journal article
Document type Scientific Article
Keywords Flow ; Flow Rate Resistance ; Hirst-type ; Pollen ; Spore ; Trap
Language english
Publication Year 2023
HGF-reported in Year 2023
ISSN (print) / ISBN 0393-5965
e-ISSN 1573-3025
Journal Aerobiologia
Quellenangaben Volume: 39, Issue: 1, Pages: 143–148 Article Number: , Supplement: ,
Publisher Pitagora Ed.
Publishing Place Bologna
Reviewing status Peer reviewed
POF-Topic(s) 30202 - Environmental Health
Research field(s) Allergy
PSP Element(s) G-505400-001
Grants COST ADOPT
Scopus ID 85148345934
Erfassungsdatum 2023-11-30