Warmbrunn, J.* ; Straube, C.* ; Haase, H.U.* ; Sinnecker, D.* ; Laugwitz, K.L.* ; Combs, S.E. ; Schneider, S.* ; Habermehl, D.*
Influence of radiotherapy on cardiac implantable devices and leads-a single-institution analysis and critical evaluation of current guidelines.
Strahlenther. Onkol. 201, 463-471 (2025)
PURPOSE: Increasing life expectancy and advances in cancer treatment will lead to more patients needing both radiation therapy (RT) and cardiac implantable electronic devices (CIEDs). CIEDs, including pacemakers and defibrillators, are essential for managing cardiac arrhythmias and heart failure. Telemetric monitoring of CIEDs checks battery status, lead function, settings, and diagnostic data, thereby identifying software deviations or damage. This study evaluates the German Society for Radiation Oncology (DEGRO)/German Society for Cardiology (DGK) guideline, assessing real-world complications and risk factors and analyzing pacemaker and implantable cardioverter-defibrillator (ICD) lead function for their predictive value concerning device malfunction. METHODS: A total of 54 patients with pacemakers or ICDs who underwent radiation therapy were identified. Demographics, treatment courses, and device information from physical and digital records were extracted. DEGRO/DGK risk groups and pacemaker dependency at the start of RT were assessed. Delineation of the devices and lead insertion sites was performed in the treatment planning system. Dosimetric information from the treatment plans was then correlated with reports of standardized device checks. RESULTS: Over 80% of patients were treated with dual-chamber pacemakers or cardiac resynchronization therapy (CRT), and 16.7% had ICDs. One third of patients were pacemaker dependent. 59.3% of patients were in the low-risk category, 29.3% in the medium-risk, and 11.1% in the high-risk category. Thoracic irradiation resulted in the highest median dose to devices. Lead parameter deviations exceeding thresholds were found in 14.8% for the stimulation threshold and 13.5% for sensing. Device malfunctions occurred in 3.7% of cases, both involving electrical resets and neutron-producing radiation (beam energy 10 megaelectron volt (MV) or higher). CONCLUSION: Collecting lead parameters in addition to secure malfunctions like electrical restarts and memory erasure revealed a significant proportion of treatment courses with temporary changes, though no correlation with individual treatment plans or adverse events was found. The focus on reducing neutron-producing radiation could be further supported.
Impact Factor
Scopus SNIP
Web of Science
Times Cited
Scopus
Cited By
Altmetric
Publication type
Article: Journal article
Document type
Scientific Article
Thesis type
Editors
Keywords
Cardiac Implantable Electronic Devices ; Device Malfunctions ; Neutron-producing Radiation ; Pacemaker Dependency ; Radiation Therapy; Degro/dgk-guideline; Latest Generation; Pacemaker; Consensus; Malfunction
Keywords plus
Language
english
Publication Year
2025
Prepublished in Year
0
HGF-reported in Year
2025
ISSN (print) / ISBN
0179-7158
e-ISSN
1439-099X
ISBN
Book Volume Title
Conference Title
Conference Date
Conference Location
Proceedings Title
Quellenangaben
Volume: 201,
Issue: 4,
Pages: 463-471
Article Number: ,
Supplement: ,
Series
Publisher
Urban & Vogel
Publishing Place
Tiergartenstrasse 17, D-69121 Heidelberg, Germany
Day of Oral Examination
0000-00-00
Advisor
Referee
Examiner
Topic
University
University place
Faculty
Publication date
0000-00-00
Application date
0000-00-00
Patent owner
Further owners
Application country
Patent priority
Reviewing status
Peer reviewed
POF-Topic(s)
30203 - Molecular Targets and Therapies
Research field(s)
Radiation Sciences
PSP Element(s)
G-501300-001
Grants
Projekt DEAL
Copyright
Erfassungsdatum
2025-03-19