Prefocusing inside a linear ion beam guide - a SIMION study.
Int. J. Mass Spectrom. 325, 25-29 (2012)
A set of 14 stacked ring electrodes was studied by the SIMION ion trajectory simulation program under a pressure of 1 Pa with different applied RF frequencies and RF voltage amplitudes when a classical DC gradient of 38 V/cm is applied. Ion–neutral collision effects were considered by adopting a successful and integrated hard sphere, elastic, ion–neutral collision user program provided by SIMION. Mass dependent ion transmission efficiencies could be obtained. In the classical operation mode of the ion beam guide, a RF frequency of 2 MHz was crucial to obtain very good ion transmission efficiency along a broad mass range. Lower RF frequencies (<1.5 MHz) limited the good transmission efficiencies only to high masses with m/z > 500 amu. SIMION can reproduce expected efficiency curves which show that low m/z ions below 150 amu represent a challenge for such linear ion beam devices even when high RF driving frequency of 2 MHz is applied. The simulations also show that a DC gradient of at least 30 V/cm is crucial to maintain good transmission efficiencies which are in agreement with previously published results. A new operation mode for running the same device is introduced for earlier ion focusing in between the second and fifth ring electrodes, prior to slight beam divergence after the focus point. This required a significant change of the applied DC voltages to alter the shape of the DC gradient curve along the central z axis. The newly identified operation mode allows enhancing the mass dependent ion transmission efficiency in a wide m/z range between 80 and 1000 amu when an RF frequency of 2 MHz is applied to the first ten ring electrodes.
Altmetric
Additional Metrics?
Edit extra informations
Login
Publication type
Article: Journal article
Document type
Scientific Article
Keywords
on beam guide; Transmission efficiency; Mass dependency; Operation mode; Ionization Mass-Spectrometry; Funnel Interface; Transmission; Pressure
ISSN (print) / ISBN
1387-3806
e-ISSN
1873-2798
Quellenangaben
Volume: 325,
Pages: 25-29
Publisher
Elsevier
Non-patent literature
Publications
Reviewing status
Peer reviewed
Institute(s)
Research Unit Analytical BioGeoChemistry (BGC)