Kramer, R.* ; Drexler, G.* ; Petoussi-Henß, N. ; Zankl, M. ; Regulla, D.F. ; Panzer, W.

Backscatter factors for mammography calculated with Monte Carlo methods.

Phys. Med. Biol. 46, 771-781 (2001)
DOI
Open Access Green as soon as Postprint is submitted to ZB.
The objective of this study is to establish a comprehensive set of backscatter factors for mammography based on the exposure model proposed by the European Protocol on Dosimetry in Mammography. The Monte Carlo calculated backscatter factors (BSFs) presented in this study are for various exposure conditions encountered in mammographic practice as well as in calibration procedures. The data demonstrate the variation of the BSF as a function of the exposure parameters, hence enabling a better match with calibration conditions and, at the same time, reviewing the BSF data already recommended by the European Protocol. Furthermore, earlier data for BSF for general diagnostic radiology are validated.
Impact Factor
Scopus SNIP
Web of Science
Times Cited
Scopus
Cited By
Altmetric
2.013
0.000
10
19
Tags
Annotations
Special Publikation
Hide on homepage

Edit extra information
Edit own tags
Private
Edit own annotation
Private
Hide on publication lists
on hompage
Mark as special
publikation
Publication type Article: Journal article
Document type Scientific Article
Keywords X-ray-spectra; semiempirical model; cross-sections
Language english
Publication Year 2001
HGF-reported in Year 0
ISSN (print) / ISBN 0031-9155
e-ISSN 1361-6560
Quellenangaben Volume: 46 , Issue: 3, Pages: 771-781 Article Number: , Supplement: ,
Publisher Institute of Physics Publishing (IOP)
Publishing Place Bristol
Reviewing status Peer reviewed
POF-Topic(s) 30504 - Mechanisms of Genetic and Environmental Influences on Health and Disease
Research field(s) Radiation Sciences
PSP Element(s) G-501100-008
Scopus ID 0035109110