Mieth, M.* ; Schubert, J.K.* ; Gröger, T.M. ; Sabel, B.* ; Kischkel, S.* ; Fuchs, P.* ; Hein, D.* ; Zimmermann, R. ; Miekisch, W.*

Automated needle trap heart-cut GC/MS and needle trap comprehensive two-dimensional GC/TOF-MS for breath gas analysis in the clinical environment.

Anal. Chem. 82, 2541-2551 (2010)
DOI PMC
Open Access Green möglich sobald Postprint bei der ZB eingereicht worden ist.
This study was intended to evaluate low-volume (20 mL) multibed needle trap (NTD) sampling combined with heart-cut gas chromatography/mass spectrometry (GC/MS) and comprehensive two-dimensional gas chromatography/time-of-flight mass spectrometry (GC x GC/TOF-MS) for trace gas analysis under clinical conditions. NTDs, high-throughput automatic desorption and separation systems, were tested in vitro and within a study in 11 patients undergoing cardiac surgery with respect to reproducibility, reliability, and clinical applicability. NTD-heart-cut GC/MS analysis of standard mixtures containing different volatile organic compounds (VOCs) yielded relative standard deviations (RSDs) from 4.0% to 18.5%. Substance adsorption was stable for 1 day if NTDs were closed on both ends and was stable for approximately 7.8 h when NTD tip ends had to be left open during autosampler storage. Even in the presence of high concentrations of contaminants linearity of heart-cut GC/MS was conserved. In patients' breath potential biomarkers could be determined even in the presence of very high concentrations of sevoflurane. Profiles of blood-borne biomarkers, intravenous drugs, and clinical contaminants were characterized. Comprehensive GC x GC/TOF-MS may be used as a screening tool for new biomarkers, if patterns are generated from deconvoluted normalized areas. Needle trap sampling in combination with hyphenated chromatographic techniques can thus be used to provide well-tailored solutions for complex problems occurring in clinical breath analysis.
Impact Factor
Scopus SNIP
Web of Science
Times Cited
Scopus
Cited By
Altmetric
5.214
2.320
109
115
Tags
Anmerkungen
Besondere Publikation
Auf Hompepage verbergern

Zusatzinfos bearbeiten
Eigene Tags bearbeiten
Privat
Eigene Anmerkung bearbeiten
Privat
Auf Publikationslisten für
Homepage nicht anzeigen
Als besondere Publikation
markieren
Publikationstyp Artikel: Journalartikel
Dokumenttyp Wissenschaftlicher Artikel
Schlagwörter Flight mass-spectrometry; X GC-tofms; Volatile organic-compounds; Sorption trap; Chromatography; Biomarkers; Devices; Samples; Flow; Chemiluminescence
Sprache englisch
Veröffentlichungsjahr 2010
HGF-Berichtsjahr 2010
ISSN (print) / ISBN 0003-2700
e-ISSN 1520-6882
Zeitschrift Analytical Chemistry
Quellenangaben Band: 82, Heft: 6, Seiten: 2541-2551 Artikelnummer: , Supplement: ,
Verlag American Chemical Society (ACS)
Begutachtungsstatus Peer reviewed
Institut(e) Institute of Ecological Chemistry (IOEC)
Cooperation Group Comprehensive Molecular Analytics (CMA)
POF Topic(s)
Forschungsfeld(er)
PSP-Element(e) G-505100-002
PubMed ID 20170082
Scopus ID 77949866490
Erfassungsdatum 2010-03-15