Flynn, S.* ; Price, T.* ; Allport, P.P.* ; Patallo, I.S.* ; Thomas, R.* ; Subiel, A.* ; Bartzsch, S. ; Treibel, F.* ; Ahmed, M.* ; Jacobs-Headspith, J.* ; Edwards, T.* ; Jones, I.* ; Cathie, D.* ; Guerrini, N.* ; Sedgwick, I.*

First demonstration of real-time in-situ dosimetry of X-ray microbeams using a large format CMOS sensor.

Nucl. Instrum. Methods Phys. Res. Sect. A 978:164395 (2020)
DOI
Open Access Green möglich sobald Postprint bei der ZB eingereicht worden ist.
Microbeam radiotherapy is a novel type of radiotherapy in which narrow beams of radiation (typically less than 500 mu m) are spatially fractionated, delivering a non-uniform distribution to the target tumour volume. Due to the very high dose gradients and very small beams involved, new dosimetric techniques are required for translation into clinical practise. Current real-time beam monitoring is typically performed using 1 dimensional silicon strip detectors or wire chambers, with 2D beam information measured offline using radiochromic film (requiring a minimum of 24 h to self-develop).Using an Xstrahl SARRP X-ray irradiation device with a bespoke microbeam collimator at the Technical University of Munich, Germany, the newly developed vM1212 detector was exposed to a variety of microbeams (220 kV, nominal slit widths 0-100 mu m) for evaluation of in vivo real time verification.The performance of the detector was assessed by changing the collimator slit width (and thus microbeam FWHM) mid-irradiation. Microbeam FWHMs of 130-190 mu m could be measured in this manner in addition to temporally monitoring other basic parameters such as the radiation intensity. More advanced parameters could be calculated as the tungsten slits within the microbeam collimator opened and closed such as the rate of change of FWHM; the peak-valley-dose-ratio (PVDR); and the sub-pixel movement of each microbeam peak.This work demonstrates the potential of radiation hard CMOS sensors in radiotherapy for in vivo real-time monitoring of X-ray microbeams FWHM, intensity and position.
Impact Factor
Scopus SNIP
Scopus
Cited By
Altmetric
1.265
1.154
1
Tags
Anmerkungen
Besondere Publikation
Auf Hompepage verbergern

Zusatzinfos bearbeiten
Eigene Tags bearbeiten
Privat
Eigene Anmerkung bearbeiten
Privat
Auf Publikationslisten für
Homepage nicht anzeigen
Als besondere Publikation
markieren
Publikationstyp Artikel: Journalartikel
Dokumenttyp Wissenschaftlicher Artikel
Schlagwörter Cmos ; Microbeam ; Radiotherapy ; Dosimetry; Radiation-therapy; 9l Gliosarcoma; Synchrotron; Beams
Sprache englisch
Veröffentlichungsjahr 2020
HGF-Berichtsjahr 2020
ISSN (print) / ISBN 0168-9002
e-ISSN 1872-9576
Quellenangaben Band: 978, Heft: , Seiten: , Artikelnummer: 164395 Supplement: ,
Verlag Elsevier
Verlagsort Radarweg 29, 1043 Nx Amsterdam, Netherlands
Begutachtungsstatus Peer reviewed
POF Topic(s) 30203 - Molecular Targets and Therapies
Forschungsfeld(er) Radiation Sciences
PSP-Element(e) G-501300-001
Förderungen UK government's Department for Business, Energy and Industrial Strategy
Scopus ID 85088323424
Erfassungsdatum 2020-10-05