Ahmed, M. ; Beyreuther, E.* ; Gantz, S.* ; Horst, F.* ; Meyer, J.* ; Pawelke, J.* ; Schmid, T.E. ; Stolz, J. ; Wilkens, J.J.* ; Bartzsch, S.
Design and dosimetric characterization of a transportable proton minibeam collimation system.
Front. Oncol. 14:1473625 (2024)
BACKGROUND: Proton Minibeam Radiation Therapy has shown to widen the therapeutic window compared to conventional radiation treatment in pre-clinical studies. The underlying biological mechanisms, however, require more research. PURPOSE: The purpose of this study was to develop and characterize a mechanical collimation setup capable of producing 250µm wide proton minibeams with a center-to-center distance of 1000µm. METHODS: To find the optimal arrangement Monte Carlo simulations were employed using the Geant4 toolkit TOPAS to maximize key parameters such as the peak-to-valley dose ratio (PVDR) and the valley dose rate. The experimental characterization of the optimized setup was carried out with film dosimetry at the University Proton Therapy beamline in Dresden and the proton beamline of the University of Washington Medical Center in Seattle with 150MeV and 50.5MeV, respectively. A microDiamond detector (PTW, Freiburg, Germany) was utilized at both beamlines for online proton minibeam dosimetry. RESULTS: A PVDR of 10 was achieved in Dresden and a PVDR of 14 in Seattle. Dosimetry measurements were carried out with EBT3 films at a depth of 5mm in a polymethylmethacrylate (PMMA) phantom. When comparing film dosimetry with the microDiamond, excellent agreement was observed in the valleys. However, the peak dose showed a discrepancy of approximately 10% in the 150MeV beam and 20% in the 50.5MeV beam between film and microDiamond. DISCUSSION: The characteristics of the minibeams generated with our system compares well with those of other collimated minibeams despite being smaller. The deviations of microDiamond measurements from film readings might be subject to the diamond detector responding differently in the peak and valley regions. Applying previously reported correction factors aligns the dose profile measured by the microDiamond with the profile acquired with EBT3 films in Dresden. CONCLUSION: The novel proton minibeam system can be operated independently of specific beamlines. It can be transported easily and hence used for inter-institutional comparative studies. The quality of the minibeams allows us to perform in vitro and in vivo experiments in the future. The microDiamond was demonstrated to have great potential for online dosimetry for proton minibeams, yet requires more research to explain the observed discrepancies.
Impact Factor
Scopus SNIP
Web of Science
Times Cited
Scopus
Cited By
Altmetric
Publikationstyp
Artikel: Journalartikel
Dokumenttyp
Wissenschaftlicher Artikel
Typ der Hochschulschrift
Herausgeber
Schlagwörter
Micro Diamond ; Monte Carlo ; Proton Minibeam Collimation ; Proton Minibeam Dosimetry ; Proton Minibeam Radiation Therapy ; Spatially Fractionated Radiation Therapy
Keywords plus
Sprache
englisch
Veröffentlichungsjahr
2024
Prepublished im Jahr
0
HGF-Berichtsjahr
2024
ISSN (print) / ISBN
2234-943X
e-ISSN
2234-943X
ISBN
Bandtitel
Konferenztitel
Konferzenzdatum
Konferenzort
Konferenzband
Quellenangaben
Band: 14,
Heft: ,
Seiten: ,
Artikelnummer: 1473625
Supplement: ,
Reihe
Verlag
Frontiers
Verlagsort
Avenue Du Tribunal Federal 34, Lausanne, Ch-1015, Switzerland
Tag d. mündl. Prüfung
0000-00-00
Betreuer
Gutachter
Prüfer
Topic
Hochschule
Hochschulort
Fakultät
Veröffentlichungsdatum
0000-00-00
Anmeldedatum
0000-00-00
Anmelder/Inhaber
weitere Inhaber
Anmeldeland
Priorität
Begutachtungsstatus
Peer reviewed
POF Topic(s)
30203 - Molecular Targets and Therapies
Forschungsfeld(er)
Radiation Sciences
PSP-Element(e)
G-501300-001
Förderungen
German Academic Exchange Service (DAAD)
Helmholtz Society, Helmholtz AI projects
Copyright
Erfassungsdatum
2025-01-10